| Verify that ambient temperature is < 90°F |
🗸 |
|
|
| Check and clean probe cover gaskets, hinges, and lid(s) of ice and snow. More frequent cleaning may be required, depending on use and environmental conditions. |
🗸 |
|
|
| Check air filter. Clean or replace as needed. |
|
|
🗸 4X |
| Vacuum condenser coils as needed. (Use caution-SHARP) |
|
🗸 |
|
| Clean unit with mild detergent, using baking soda and water to treat odors. |
🗸 |
|
|
| Verify the unit alarms are active and limits are properly set for the content. |
|
🗸 |
|
| Check alarm back-up battery, replace as necessary. |
🗸 |
|
|
| Check chart recorder backup battery (9V), replace as necessary. |
🗸 |
|
|
| Inspect gaskets for tears and replace if compromised. Clean with mild detergent, check seal with dollar/paper test. |
|
🗸 |
|
| Vacuum condenser condensate pan. Use caution as most pans have a heater for evaporation- may be hot. |
|
🗸 |
|
| Defrost freezer at 3/8" frost thickness. Must be turned off for 24 hours before restarting. |
|
|
🗸 |