Custom Biogenic
Cleaning & Maintenance
Cleaning Exterior
To clean the exterior, use a mild, non-abrasive household type cleaner mixed with warm water. Apply by using a damp cloth that you soaked in the cleaning solution and wrung out. Do not expose LCD Display to liquids as this may cause damage.
Sanitizing and Decontamination of Interior
- The best decontamination method is a mixture of 30% chlorine bleach with 70% water solution.
- Other methods such as any household detergent, mild soap solution, bleach, hydrogen peroxide, and denature alcohol can be safely used.
- Do not use any petroleum based cleaning solution.
- To perform this sanitizing procedure, cover all inner surfaces with solution, let stand for 30 minutes, and remove.
- Spraying the solution into the inner vessel is preferred.
- Rinse the decontaminated surfaces with clean water and ensure all cleaner residue has been removed
- Allow the tank to dry completely before putting back into service.
If you any questions concerning the decontamination process, please contact Custom BioGenics Systems Technical Support 1-800-523-0072.
System Check
- Test all alarm functions for proper operation.
- Check any connected Remote Alarms or Automatic Dialing Systems to ensure proper operation.
- Check for leaks at all connection points of the liquid nitrogen lines.
Solenoid Valve Maintenance
Alert
This should be done only as required, and is not a necessary part of maintenance.
- The solenoid valves may be opened and cleaned if they are not working smoothly. If this is done, use the diagram below to re-assemble the valves properly.
- Wipe seals with a clean, lint-free cloth. Distilled ammonia may be used to wipe seals if needed.
- If valve is found to be defective, the entire valve must be replaced, individual parts are not available.
- Valve Body
- Plunger
- Flange
- Valve Coil Locknut
- Coil
- Spec. Plate
- Valve Cap