Cleaning & Maintenance

Cleaning Exterior

To clean the exterior, use a mild, non-abrasive household type cleaner mixed with warm water. Apply by using a damp cloth that you soaked in the cleaning solution and wrung out. Do not expose LCD Display to liquids as this may cause damage.

Sanitizing and Decontamination of Interior

  • The best decontamination method is a mixture of 30% chlorine bleach with 70% water solution.
  • Other methods such as any household detergent, mild soap solution, bleach, hydrogen peroxide, and denature alcohol can be safely used.
  • Do not use any petroleum based cleaning solution.
  • To perform this sanitizing procedure, cover all inner surfaces with solution, let stand for 30 minutes, and remove.
  • Spraying the solution into the inner vessel is preferred.
  • Rinse the decontaminated surfaces with clean water and ensure all cleaner residue has been removed
  • Allow the tank to dry completely before putting back into service.

If you any questions concerning the decontamination process, please contact Custom BioGenics Systems Technical Support 1-800-523-0072.

System Check

  1. Test all alarm functions for proper operation.
  2. Check any connected Remote Alarms or Automatic Dialing Systems to ensure proper operation.
  3. Check for leaks at all connection points of the liquid nitrogen lines.

Solenoid Valve Maintenance

Alert

This should be done only as required, and is not a necessary part of maintenance. 

  • The solenoid valves may be opened and cleaned if they are not working smoothly. If this is done, use the diagram below to re-assemble the valves properly.  
  • Wipe seals with a clean, lint-free cloth. Distilled ammonia may be used to wipe seals if needed.
  • If valve is found to be defective, the entire valve must be replaced, individual parts are not available.
  1. Valve Body
  2. Plunger
  3. Flange
  4. Valve Coil Locknut
  5. Coil
  6. Spec. Plate
  7. Valve Cap
Solenoid Valve